How to handle incomplete or interrupted tests?

Before the incubation has started:

When starting a test on the DAS the user can reset the data inserted (operator, layout definition, barcode, method) by pressing the Stop button on the DAS. (See also question 'How can I start a test?')

During incubation:

Once the incubation has started (Running status) and till just before the incubation is finished, the running measurement can be aborted by long pressing the Stop button (at least 10 seconds).

The system will be back to Idle.

2) Interrupted tests

  • If the DAS automatically aborts the incubation within 10 seconds from the start, check that the partial plate position on the grid matches the layout selection.

See below instructions from "How can I start a test?"

Note: In case of partial plate incubation with software 0.1.99 or higher, an additional check is performed to confirm that the block(s) has been placed in the correct grid position. When the partial plate position and/or size does not match the layout selection, the measurement will be aborted within the first 10 seconds of the run and the cups will be classified as “Not Active (NA)”. This function allows the user to reposition the plate as indicated in the “prepare” step 6.

6. Confirm the plate position on the grid - Deselect unused blocks

On the layout table, select the number and position of blocks to incubate. By default, the full plate is selected on the display. For block or partial plate incubation, select only the position(s) were the block(s) are placed. The incubation display is 180° rotated in comparison to the DAS heating grid. A01 well is located on the lower right corner of the plate. An example of a single block incubation on the right side of the grid is shown in Figure below:

  • If the measurement is aborted at


? Got any questions? Need help? Get in touch support@delvotest.com

Was this helpful?
Nam quis tincidunt porttitor turpis eu semper. Netus sed pellentesque proin nec sit cursus quis ut leo.
Important
Although diligent care has been used to ensure that the information provided herein is accurate, nothing contained herein can be construed to imply any representation or warranty for which we assume legal responsibility, including without limitation any warranties as to the accuracy, currency or completeness of this information or of noninfringement of third party intellectual property rights. The content of this document is subject to change without further notice. This document is non-controlled and will not be automatically replaced when changed. Please contact us for the latest version of this document or for further information. Since the user's product formulations, specific use applications and conditions of use are beyond our control, we make no warranty or representation regarding the results which may be obtained by the user. It shall be the responsibility of the user to determine the suitability of our products for the user's specific purposes and the legal status for the user's intended use of our products.©DSM Food Specialties B.V. | Trade Register Number 27235314